6
T3
Channel 3 output -- same specifications as Channel 1.
T4 Output
Channel 4 output -- same specifications as Channel 1.
Ext/Gate
External Trigger or Gate signal input -- used as either an external signal to trigger the genera-
tion of pulses or as a gate signal to enable and disable the outputs.
Back Panel
Output Voltage Adjustment
The maximum output voltage of all output channels is adjusted by turning the potentiometer la-
beled VOUT. Turn clockwise to increase the voltage, counter-clockwise to decrease the volt-
age. The maximum voltage possible is 20 Volts into a high-impedance load or 10 Volts into a
50 Ohm load.
Serial Port
This is a 9-pin D-sub connector for the RS232 interface that comes standard with the 9314E.
It is pinned to be directly connected to a standard IBM-PC serial port -- use a straight-through
cable (not a null modem cable) to connect to a PC.
GPIB Port
A GPIB (IEEE 488) interface is available as an option with the 9314E.
AC Input
A standard IEC instrument-type cord should be used. The 9314E is supplied with a power
cord that fits the type of socket used in your country.
Before plugging in, please make sure that the voltage selection switch is set
to the proper setting for your country (either 110V or 220V).